Wearable-Grade Lead Reduction Disproportionately Degrades ECG AI Performance in Elderly Patients: Evidence from PTB-XL and MIT-BIH

摘要 / Abstract

Consumer wearable devices increasingly use single-lead electrocardiograms (ECGs) for cardiac monitoring, but these signals contain substantially less spatial information than the clinical 12-lead standard. Whether this reduction dispro- portionately affects older adults, who often present with more complex cardiac conditions, remains poorly understood. In this study, we evaluated the impact of lead reduction on AI-ECG diagnostic performance across age groups. A 1D resid- ual neural network was trained on 21,091 PTB-XL ECG recordings spanning five diagnostic superclasses and assessed using 12-, 6-, 2-, and 1-lead configurations. Under the full 12-lead setting, model accuracy declined from 84.5% in patients younger than 40 years to 66.2% in patients aged 75 years or older. Progressive lead reduction further widened this gap. Under the 1-lead configuration, accuracy decreased by 14.1 percentage points in the 75+ group but by only 0.4 percent- age points in the