medRxiv (Medicine)
2026-06-11 00:00
DOI:
HASH:9ab712c25b57181efc962290cd2be2f5
Assessment of occupational aerosol exposure for laboratory technicians: A quantitative study using {Phi}X174 phage as a substitute virus
作者:
摘要 / Abstract
This study aimed to clarify aerosol exposure risks throughout the workflow of a Biosafety Level 2 (BSL-2) polymerase chain reaction (PCR) laboratory, validate the suitability of the {Phi}X174 bacteriophage as an indicator virus, and provide evidence for biosafety control measures. The {Phi}X174 bacteriophage was used to simulate viral samples, and a concentration-bacteriophage plaque standard curve was constructed (R2=0.998). Five operational steps in a simulated PCR laboratory were quantitatively monitored for aerosol concentration using double-layer agar plates, with blank controls used to eliminate interference. Statistical analysis was employed to identify risk differences. Sample homogenization ((5.67 {+/-} 1.23) x 104 plaque-forming units (PFU)/m3) and nucleic acid extraction ((3.45 {+/-} 0.89) x 104 PFU/m3) were identified as high-/very high-risk steps. The viral load in the samples was strongly positively correlated with the aerosol concentration (r = 0.926, P